Autoclave Testing

Autoclave users can now have the effectiveness of their disinfection process verified. This service (upon request) is provided by EHS. Steam autoclaves (250F gravity or 270F pre-vacuum) are tested by using ATI Biological Indicator Kits. These kits consist of a self contained packet that is cycled through the autoclave checking both temperature and sterilization.

Testing Procedure

  1. Request an autoclave test by contacting Robert Golden.
  2. EHS delivers the test pack with instructions.
  3. The test pack is cycled through the autoclave by the on site employee.
  4. EHS picks up the test pack with-in two hours after cycling.
  5. The pack is opened by EHS and the steam temperature indicator results are noted.
  6. The biological indicator vial is placed in the incubator and checked for growth after 24 and 48 hours.
  7. Results are sent to the contact person at the location.
  8. Unsatisfactory results will need to be followed-up by the users. This is done by contacting the Physical Facilities Work Control Center, (765) 494-9999, or by contacting the autoclave service provider.

If you have any questions concerning autoclave testing contact Robert Golden.

Campus Safety Contacts

call 911

If you see something, say something.

Purdue Police
Phone: (765) 494-8221

Purdue Fire
Phone: (765) 494-6919

REPORT A SAFETY CONCERN

Sign up for Emergency Text Messages

Do you want to recognize one of our staff for a job well done? Nominate the staff member(s) for a Bravo Award here!

Purdue University, 610 Purdue Mall, West Lafayette, IN, 47907, 765-494-4600

© 2019 Purdue University | An equal access/equal opportunity university | Integrity Statement | Copyright Complaints | Brand Toolkit | Maintained by Administrative Operations

Contact Administrative Operations for accessibility issues with this page | Accessibility Resources | Contact Us