Time-Sensitive Chemicals


Introduction

Many chemicals have a shelf-life and can become more hazardous as time goes on. This webpage discusses two classes of time-sensitive hazardous chemicals: peroxide forming solvents and certain compressed gas cylinders. It is Extremely important that these chemicals are identified and managed properly. If any time-sensitive chemicals are found to be past the manufacturer’s expiration date, they must be submitted to EHS for hazardous waste disposal immediately.

It is important to note that the time-sensitive chemicals listed on this webpage are not comprehensive. Researchers should always review the material safety data sheets for information regarding time-sensitive chemicals and expiration dates.


Peroxide Forming Solvents

Autoxidation in common laboratory solvents can lead to unstable and potentially explosive peroxide formation. The reaction can be initiated by exposure to air, heat, light, or contaminants. Most of these solvents are available with inhibitors to slow the peroxide formation. Examples of inhibitors include BHT (2,6-di-tert-butyl-4-methyl phenol) and Hydroquinone. There are three categories of peroxide formers:

  1. Group A chemicals are those which form explosive levels of peroxides after prolonged storage, especially after exposure to air without concentration. Test these for peroxide formation before using and discard 3 months after opening. 
Butadiene Divinylacetylene Tetrafluoroethylene
Chloroprene Isopropyl ether Vinylidene chloride

  1. Group B chemicals form peroxides that are hazardous only on concentration by distillation or evaporation. Test these before distillation and discard after 12 months. 
Acetal Dicyclopentadiene Methyl isobutyl ketone
Acetaldehyde Diethyl ether 4-Methyl-2-pentanol
Benzyl alcohol Diethylene glycol dimethyl ether 2-Pentanol
2-Butanol Dioxane 4-Penten-1-ol
Cumene Ethylene glycol dimethyl ether 1-Phenylethanol
Cyclohexanol 4-Heptanol 2-Phenylethanol
2-cyclohexen-1-ol 2-Hexanol 2-Propanol
Cyclohexene Methylacetylene Tetrahydrofuran
Decahydronaphthalene 3-Methyl-1-butanol Tetrahydronaphthalene
Diacetylene Methylcyclopentane Vinyl ether

  1. Group C chemicals consists of monomers which form peroxides that can initiate explosive polymerization. Inhibited monomers should be tested before use and discarded after 12 months. Uninhibited monomers should be discarded 24 hours after opening.
Acrylic acid Methyl methacrylate Vinyl chloride
Acrylonitrile Styrene Vinyl pyridine
Butadiene Tetrafluoroethylene Vinyladiene chloride
Chloroprene Vinyl acetate  
Chlorotrifluoroethylene Vinyl acetylene  

General Guidelines

  • Solvents containing inhibitors should be used whenever possible.
  • All peroxide forming solvents should be tested prior to distillation.
  • Peroxide forming solvents should be purchased in limited quantities.
  • Peroxide forming solvents should be marked with the purchase date and the date opened.
  • Peroxide forming solvents should be sealed tightly and stored away from light and heat.
  • Periodic testing should be done on opened containers and the results marked on the containers.

Testing

  • Obtain test strips for the range of 0-100 ppm peroxide
  • Record the test results on the bottle
  • If the test results are 100 ppm or greater, arrange for EHS disposal

Time-Sensitive Compressed Gas Cylinders

  • Hydrogen Fluoride, Anhydrous
  • Hydrogen Bromide, Anhydrous
  • Hydrogen Sulfide, Anhydrous
  • Hydrogen Cyanide, Anhydrous
  • Hydrogen Chloride, Anhydrous

The compressed gases listed above have a shelf-life provided by the manufacturer that must be strictly followed. There have been numerous incidents involving these compounds related to storage past the expiration date. For example, Hydrogen fluoride (HF) and Hydrogen bromide (HBr) cylinders have a shelf-life of one to two years, depending on the vendor. Over time, moisture can slowly enter the cylinder, which initiates corrosion. As the corrosion continues, HF and/or HBr slowly react with the internal metal walls of the cylinder to produce Hydrogen. The walls of the cylinder weaken due to the corrosion, while at the same time the internal pressure increases due to the Hydrogen generation. Ultimately, these cylinders fail and create Extremely dangerous projectiles and a toxic gas release (Figure 1).

General Guidelines

  • Acquire gases in the smallest quantity and concentration necessary.
  • Record the date the compressed gas was received. The shelf-life vendor recommendation must be strictly followed. Once a compressed gas cylinder has exceeded its expiration date, arrange disposal through EHS.
  • Routinely inspect gas cylinders for evidence of deterioration or elevated pressure.

30-year old HF cylinder that exploded at Purdue in 2011

Figure 1: 30-year old HF cylinder that exploded at Purdue in 2011


References

  • Chemical Health and Safety, Vol. No. 3, No. 5, “Review of Safety Guidelines for Peroxidizable Organic Chemicals", September/October 1996, pp. 28-36
Campus Safety Contacts

call 911

If you see something, say something.

Purdue Police
Phone: (765) 494-8221

Purdue Fire
Phone: (765) 494-6919

REPORT A SAFETY CONCERN

Sign up for Emergency Text Messages

Do you want to recognize one of our staff for a job well done? Nominate the staff member(s) for a Bravo Award here!

Purdue University, 610 Purdue Mall, West Lafayette, IN, 47907, 765-494-4600

© 2019 Purdue University | An equal access/equal opportunity university | Integrity Statement | Copyright Complaints | Brand Toolkit | Maintained by Administrative Operations

Contact Administrative Operations for accessibility issues with this page | Accessibility Resources | Contact Us