Ultraspec 2100 PRO Optical Filter Spectrophotometer

Printable version

General Information

FIC: James Leary

Owner: Lisa Reece

Location: 2087

System Information

General Description

UV/VIS spectrophotometer modified for optical filter analyses.

There are routines for wavelength scanning and equation entry for multi-wavelength applications.

Capabilities

This standard UV/VIS spec has been modified to scan optical filters at 45 degrees and 90 degrees to incident light. Optical filter size accepted is 25 mm diameter.

  • Bandwidth: (2.0 - 3.0nm) 2.9
  • Wavelength Accuracy: 881.9nm (± 1 nm) 881.9
  • Absorbance Accuracy:
    • 220nm (1.763 - 1.781A) 1.772
    • 340nm (1.633 - 1.665A) 1.649
    • 500nm (1.477 - 1.491A) 1.484
  • Stray Light: 220 nm (<0.05%) 0.021

Materials Compatibility

Misc.

Useful Links

* Must use iLab and be a trained user to reserve a slot on this system.